To clean and disinfect the Medescan NE-MO1L Ultrasonic Mesh Nebuliser, follow these steps:
1. Remove the medication cup and accessory from the device body.
2. Open the cup lid and discard any residual medication.
3. Add 75% ethanol solution in the medication cup, cover the cup lid, and let it sit for at least 10 minutes. Gently shake it for better disinfection.
4. Immerse the accessories to be disinfected into a container with ethanol solution and close the container. Soak them in 75% ethanol solution for 10 minutes or longer.
5. Discard the disinfectant in the medication cup and rinse the cup and accessories with clear water repeatedly.
6. Fill the medication cup with clear water, assemble it to the device body, and let the device run for 10 minutes to clean the nebulizing diaphragm.
7. After cleaning, use new medical gauze to wipe away the water and fully dry the device.
8. Use 75% medicinal alcohol to wipe the surface of the device body, then air-dry or wipe-dry with a clean, soft cloth.
9. Store the device body, medication cup, and accessories in a dry, clean place.
Note: Do not disinfect the medication cup and accessories in boiling water or put them in a microwave oven for drying. The parts disinfected with disinfectant must be fully cleaned, and any residual disinfectant should be removed.
For more detailed instructions and precautions, please refer to the Medescan NE-MO1L Ultrasonic Mesh Nebuliser Instruction Manual.