To calibrate the PCE-BSK Series Precision Balance, follow these steps:
1. Ensure that the balance has been exposed to stable environmental conditions for at least 30 minutes before calibration.
2. In weighing mode, press and hold the "CAL" key for approximately 4 seconds. The display should show "I.CAL Ig".
3. Press the ">0<" key to display a weight value for calibration.
4. Press the ">0<" key to apply the weight value, or press the counting key and the CAL key to make changes, then confirm with ">0<".
5. Apply the weight that is displayed on the screen. A line will appear, followed by the non-flashing weight.
6. Remove the weight, and a line will be displayed, followed by "0.00 g". The calibration is now completed.
Note: For a three-point calibration, follow the same steps but select "2.CAL 3" in step 2 and apply three different weights instead of one.
It is important to calibrate the balance regularly to ensure accurate measurements.