To adjust or calibrate the PCE-MA Instruments Precision Balance, follow these steps:
1. Make sure the balance is properly set up and leveled.
2. Press and hold the power button for approximately 6 seconds until "-CAL-" appears on the display.
3. After "-CAL-" appears, "100.00 g" will start flashing.
4. Place a 100 g calibration weight in the middle of the weighing platform.
5. The display will change from "100.00 g" to "=====".
6. When the display changes back to "100.00 g", remove the calibration weight from the weighing platform.
7. The display will show "=====" again and then change to "0.00 g" automatically.
8. The adjustment or calibration of the balance is now complete.
Please note that it is important to position the balance in a stable and vibration-free environment before performing the adjustment. Additionally, the adjustment should only be done after the balance has been allowed to warm up for approximately 30 minutes.
For more detailed instructions and safety precautions, please refer to the PCE-MA Instruments Precision Balance User Manual.