Answer

Mar 29, 2025 - 01:05 PM
To adjust the output voltage on the Velleman LABPS5005 Lab Power Supply, follow these steps:
1. Switch on the device.
2. Turn the current adjustment xxxx fully clockwise to the maximum.
3. Turn the voltage adjustment xxxx to obtain a voltage that is close to the desired value. Use the fine-tune xxxx if applicable. The constant voltage (C.V.) indicator will light up on the display.
To adjust the output current on the Velleman LABPS5005 Lab Power Supply, follow these steps:
1. Switch on the device.
2. Turn both the current adjustment knobs fully anticlockwise to the minimum and the voltage adjustment knobs fully clockwise to the maximum.
3. Short-circuit the output terminals by interconnecting them with the appropriate cable (60 V/16 A DC cable).
4. Turn the current adjustment xxxx to the required current-limiting value. Use the fine-tune xxxx if applicable. The constant current (C.C.) indicator will light up on the display.
5. Remove the short-circuit cable from both output terminals.
6. Turn the voltage adjustment xxxx to the required value. Use the fine-tune xxxx if applicable. The C.V. indicator will light up on the display.
7. Connect the load to the output terminals. If the current exceeds the current-limiting point, the C.C. indicator will light up on the display.
Please refer to the Velleman LABPS5005 Lab Power Supply user manual for more detailed instructions and illustrations.