Answer

Mar 26, 2025 - 08:27 AM
To calibrate the PeakTech 5220 Coating Thickness Gauge, you have three options: basic calibration, zero calibration, and multi-point calibration. Here are the steps for each calibration method:
1. Basic Calibration:
- Ensure that you have a smooth calibration sample with the same curvature radius and substrate as the sample you will be measuring.
- Follow the steps in the manual (Chapter 4.1) to perform basic calibration.
2. Zero Calibration:
- In measure mode, press the "CAL" key once to enter calibration mode.
- Place the probe on an uncoated sample and raise it after one beep. The LCD will show "X".
- Press and hold the "ZERO" key until you hear a series of beeps. The LCD will show "0" and the "ZERO" symbol.
- Repeat the previous step several times to obtain a mean value.
- Press the "CAL" key to exit calibration mode.
3. Multi-point Calibration:
- In measure mode, press the "CAL" key once to enter calibration mode.
- Perform zero calibration according to the steps mentioned above.
- Lay the calibration foil on an uncoated sample.
- Apply the probe and raise it after one beep. A reading will be shown on the LCD.
- Use the "Up" or "Down" key to adjust the reading to the required foil thickness value.
- Press the "Left" key to confirm and end the current calibration point.
- Repeat the previous steps for additional calibration points using different calibration foils.
- Press the "CAL" key to exit calibration mode.
Note: Each group (DIR or GEN) has individual zero calibration and multi-point calibration. It is recommended to work in single measure mode when calibrating.
Please refer to the manual (Chapter 4) for more detailed instructions and considerations for calibration.
Add New Comment