Answer

Mar 18, 2025 - 03:14 AM
To calibrate the INESA PH410T pH Meter, follow these steps:
1. Press the power key to turn on the instrument.
2. Select the measurement method. The default method is "No.001 direct reading measurement method", but you can select a different method if needed.
3. Enter the measurement method management and select the method in the method list or create a new method as required.
4. Go to the calibration settings.
5. Set the parameters, such as pH.
6. Select the standard solution group, such as NIST pH 4.01, pH 7.00, and pH 10.01.
7. Set the instrument to automatic recognition.
8. Enter the pH electrode calibration by selecting "Calibrate" and then "pH Calibration".
9. Place the cleaned electrode into the pH 4.01 standard solution and wait for the instrument to display "Auto Mode Matched". Once the reading is stable, press "Start".
10. If you only need 1-point calibration, after completing the calibration, press the "