Answer

Mar 06, 2025 - 07:00 PM
To use the INFICON ELT3000 Lithium Ion Battery Leak Detector, follow these steps:
1. Remove the calibration leak from the packaging and store it for 24 hours with the opening facing downwards.
2. Place the calibration leak, opening downwards, in the test chamber of the battery leak detector.
3. Evacuate the test chamber to create a vacuum.
4. The escaping solvent in gaseous form will be detected by the battery leak detector.
5. The calibration leak is used to calibrate the battery leak detector and is typically done once a day.
6. After 12 months from the manufacture date, the calibration leak must no longer be used and should be sent in for a refill.
7. Store the calibration leak in a well-ventilated place, as a closed container is not suitable.
8. Adhere to the restrictions of use and the technical specifications provided in the manual.
Please note that this is a general overview of the usage instructions. For more detailed instructions and safety precautions, refer to the INFICON ELT3000 Lithium Ion Battery Leak Detector Instruction Manual.