To replace the cable assembly on the Tektronix P6011 Passive Probe, follow these steps:
1. Unscrew the BNC plug using two 7/16 inch wrenches.
2. Unsolder the contact pin.
3. Remove the spacer insulator.
4. Unscrew the probe body and sleeve assembly from the other end of the cable assembly.
5. Unsolder the male plug from the cable.
6. Install the new cable assembly by reversing the above procedure.
Please refer to Section 2 of the Tektronix P6011 Passive Probe Instruction Manual for more details and illustrations on replacing the cable assembly.