Answer

Mar 03, 2025 - 08:46 PM
To set the stirring speed on the FP4 Portable Flocculator, use the "Stirrer rpm" xxxx on the frontal panel. The xxxx has an analogical scale around it with values of 0, 20, 40, 50, 100, and 200 rpm. Simply turn the xxxx to select the desired speed.
To set the stirring time, use the timer xxxx on the frontal panel. By turning the xxxx counterclockwise to the manual modality, the instrument will run continuously. By turning the xxxx clockwise, you can select the time of running up to 30 minutes. The stirring will automatically stop when the set time finishes.
After completing your work cycle, set the stirring speed xxxx to "0" rpm, the timer xxxx to "OFF," and turn off the light of the central column. If the instrument is not used for a long period, it is recommended to disconnect the external power supply from the plug.
For more detailed instructions and information, please refer to the VELP SCIENTIFICA F105A0117 FP4 Portable Flocculator Instruction Manual.