Answer

Feb 26, 2025 - 04:22 AM
To calibrate the NOYAFA MS12C Carbon Dioxide Detector, follow these steps:
1. Find a well-ventilated area with clean air to ensure accurate calibration. Avoid places with alcohol or perfume smells.
2. Place the instrument in the chosen location and let it sit for 5 minutes to stabilize.
3. Press and hold the brightness adjustment button for 5 seconds to initiate the calibration process. The instrument will emit a long sound.
4. After 10 seconds, the calibration will be successful, and the instrument will emit another long chirp.
Note: If the air quality is not guaranteed to be clean, find a place with good air quality and recalibrate the instrument.
It is important to follow these calibration steps to ensure accurate readings from the detector
Add New Comment