To calibrate the Linshang LS172 Colorimeter, follow these steps:
1. Power on the colorimeter by short pressing the power button.
2. Enter the "calibration" interface.
3. You can choose to perform the calibration operation as shown in the calibration animation or skip the calibration.
4. It is recommended to calibrate the colorimeter after a long period of non-usage.
5. Make sure the bottom cap is closed properly.
6. Clean the standard tile on the bottom cap if necessary.
7. If the calibration fails, it may be due to the attenuation of the light source. In this case, the colorimeter needs to be returned to the factory for inspection and repair.
Please note that the manual recommends calibrating the instrument once each year. If you have any further questions or issues with calibration, you can contact the manufacturer's service hotline or email for assistance.