Answer

Feb 26, 2025 - 02:58 AM
According to the user guide, before using the Co-Dx VZDC-R-001 Vector Smart Zika Dengue and Chikungunya RUO kit, it is recommended to clean the work surfaces in the Clean area. This can be done by following these steps:
1. Clean the work surfaces with water.
2. Use a solution of bleach (at least 10%) and water to disinfect the surfaces.
3. Rinse the surfaces with water again.
4. Finally, clean the surfaces with 70% alcohol.
It is important to follow this cleaning procedure in the specified order to ensure proper disinfection and degradation of nucleic acids on the work surfaces. Alternatively, you can use another equivalent method of cleaning that achieves the same level of disinfection and nucleic acid degradation.
By following these cleaning instructions, you can ensure a clean and suitable environment for working with the Co-Dx VZDC-R-001 Vector Smart Zika Dengue and Chikungunya RUO kit.