To perform camera spectral sensitivity measurements using the Image Engineering TE292 with iQ-LED, please follow these steps:
1. Remove the packaging material and insert the TE292 or TE292 VIS-IR into the test chart holder of the LE7 or LE7 VIS-IR illumination device.
2. Power up the LE7 or LE7 VIS-IR and connect the USB cable between the device and the host computer.
3. Using the iQ-LED software, create an E illuminant at near full power and save it to the device as described in the iQ-LED user manual.
4. Disconnect the device's USB cable.
5. Connect the EX2 spectrometer calibrator USB cable to the host computer.
6. From the control panel on the device, select the E illuminant that was saved to the device earlier.
7. Launch the camSPECS software and perform the calibration as described in the camSPECS user manual.
8. To perform camera spectral sensitivity measurements, proceed as described. The camera exposure will depart from the typical settings suggested and should be optimized as described in the camSPECS user manual.
Please refer to the user manuals for the LE7 or LE7 VIS-IR, iQ-LED software, and camSPECS software for more detailed operating instructions.