To clean the HoverMatt Air Transfer System, wipe it down with a cleaning solution used by your hospital for medical equipment disinfection. You can also use a 10:1 bleach solution (10 parts water: one part bleach) or disinfectant wipes. Follow the cleaning solution manufacturer's instructions for use, including dwell time and saturation. If the HoverMatt becomes badly soiled, it can be laundered in a washing machine with a maximum water temperature of 160°F (71°C). You can use a 10:1 bleach solution during the wash cycle. Air drying is recommended, but if using a dryer, set it to the coolest setting. Drying temperature should not exceed 115°F (46°C). The Double-Coated HoverMatt should not be put in the dryer. For preventive maintenance, perform a visual inspection before use to ensure there is no visible damage. If any damage is found, remove the HoverMatt from use and contact HoverTech International for repair. The Single-Patient Use HoverMatt is not intended to be laundered.