To clean the heating platen of the Labnet PS1000 AccuSeal Semi Automated Plate Sealer, follow these steps:
1. Open the drawer of the unit and power it down by switching it off and unplugging it from the power source.
2. Allow the unit to cool for at least 1 hour before attempting to clean the platen.
3. Using a Phillips head screwdriver, remove the retaining xxxx for the access door located on the back of the unit.
4. Remove the access door to gain access to the heating platen.
5. Clean the platen by hand using a plastic putty knife. You may use alcohol to remove any residue. Do not use a metal tool, such as a screwdriver, as it could damage the platen surface.
6. Once the platen is clean, replace the access door and secure it with the xxxx .
Please note that it is important to turn off and unplug the unit before cleaning the platen to ensure safety. Additionally, allow the platen to cool down sufficiently to avoid burns or injuries.