To install the Olink custom recipe for the NextSeq 550 sequencing run, follow these steps:
1. Unzip and place the Olink custom recipe file, "Olink_NSQ550_HighOutput_V1," in the following folder of the NextSeq 550 instrument: C:\Program Files\Illumina\NextSeq Control Software\Recipe\Custom\High\.
2. Under System Customization > Manage Instrument on the NextSeq 550, enable Custom Recipes. If this option is not selected, the custom recipe will not appear during the run setup.
Note:
- In the NCS 4.0 software version, the option to select a custom recipe will only appear after the reagent cartridge is loaded, not in the earlier setup page.
- The run must be set in manual mode to allow custom recipes.
Please refer to the Olink NextSeq 550 Explore Sequencing User Manual, specifically the "Install Olink custom recipe" section, for more detailed instructions and illustrations.