Answer

Feb 22, 2025 - 04:57 AM
To perform the MACHEREY-NAGEL NANOCOLOR Potassium 50 Tube test, follow these steps:
1. Open a test tube from the set.
2. Add 200 μL of reagent R2 to the test tube.
3. Using a digital piston pipette, add 2 mL of your sample into the test tube.
4. Seal the test tube and shake it vigorously.
5. Wait for 1 minute.
6. Add 1 level measuring spoon of reagent R3 to the test tube.
7. Seal the test tube again and shake it for 10 seconds.
8. Wait for 2 minutes.
9. Clean the outside of the test tube.
10. Place the test tube in a MACHEREY-NAGEL photometer.
11. Measure the absorbance at a wavelength of 690 nm.
Please note that when using other photometers, ensure that measurements are possible in test tubes with a 16 mm outer diameter and calibrate the method if necessary. It is also important to filter out any turbidity before performing the measurement.
For more information on safety precautions and detailed instructions, please refer to the product's label and safety data sheet, which can be downloaded from the MACHEREY-NAGEL website at www.mn-net.com/SDS.
If you have any further questions or concerns, please contact MACHEREY-NAGEL directly through their provided contact information.