Answer

Feb 15, 2025 - 03:10 AM
To assemble the Safe Aspiration Station and its accessories, follow these steps:
1. Firmly xxxx the lid onto the 4L Polypropylene (PP) bottle.
2. Place the bottle on the base unit of the Safe Aspiration Station.
3. Connect the 800 mm tubing to the lid of the bottle and to the membrane filter at the back of the station.
4. Connect the 1500 mm tubing to the lid of the bottle and insert the aspiration hand operator on the other end of the tubing.
5. Connect the sensor cable to the lid of the bottle and the back of the aspiration station.
Ensure that the lid is firmly screwed on, the release valve is closed, the hand operator is fully inserted in the tubing, and the tubings and sensor cable are correctly connected.
For a visual reference, you can refer to the "Installing" section in the user guide, which provides illustrations and detailed instructions.
Please note that proper assembly is crucial for the safe and effective operation of the Safe Aspiration Station.