Installation of the HACH CL17sc Ultra Low Range Colorimetric Chlorine Analyser involves the following steps:
1. Mounting: The analyser is designed to be wall-mounted using four ¼-inch screws. Ensure that there is adequate clearance for plumbing and controller connection.
2. Connections: Connect the sample inlet to a ¼-inch quick-connect fitting and the drain to a ½-inch ID flexible hose. The analyser receives power from connection to a Hach SC controller.
3. Controller: A Hach SC controller is required for operation and is sold separately. Connect the analyser to the controller using the provided cable, which is approximately 1.8 meters in length.
4. Reagents: The analyser is shipped with one month of reagents. The user selects either a standpipe installation kit or a pressure regulator installation kit at the time of purchase.
For more detailed instructions and illustrations, please refer to the HACH CL17sc Ultra Low Range Colorimetric Chlorine Analyser Instruction Manual.
Note: It is recommended to consult a professional or contact Hach's customer service for assistance with installation to ensure proper setup and operation of the analyser.