To connect the Oxylitre S8 Series Injector Suction Unit to a gas supply, please follow these steps:
1. Before connecting the equipment to a Gas Supply, check the unit for any visual damage and ensure that the gas control valve lever is at right angles to the white gas inlet hose.
2. Connect the Vacuum Tube to the Injector Suction Unit’s outlet and the other end to the “Vacuum” input on the Receiver Jar Lid. Fit an in-line Filter in-between if necessary (on Oxylitre 1800 ml Jars an additional Male Connector is required Part No: S7525).
3. Connect a Catheter Connecting Tube (Part No: 180FFM) to the “Patient” output connector on the Receiver Jar and connect the required Catheter to the male end of the tube. (If required, use Anti-frothing agent as per the manufacturer’s instructions).
4. Connect the Unit to a gas supply that complies with AS 2902-2005 Sleeve Index System.
Please refer to the instruction manual for illustrations and further details on the gas supply connections and recommended accessories.