Answer

Feb 07, 2025 - 12:02 AM
The Hioki LR8432 heat flow logger can connect up to 10 sensors. In the case of using Hukseflux FHF04 sensors, a total of 5 FHF04 sensors can be connected directly to the Hioki LR8432. The heat flux in W/m2 is calculated by dividing the heat flux sensor's output voltage by its sensitivity, which can be programmed directly into the datalogger. Please refer to the specifications in Table 1 for more information on the compatibility between the Hioki LR8432 and Hukseflux FHF04 sensors.