To clean the FlowTrax FT-2 Flow, Volume Testing device, please follow these instructions:
1. Press the FUNCTION key, then select the CLEAN GLASS soft key.
2. Gently pull the black tab at the top of the Glass Tube to remove it from the unit.
3. Using the provided cleaning solution and brush, vigorously scrub the inside of each tube for a minimum of 30 times.
4. Rinse the Glass Tube using a Flush Bottle filled with approved fluid.
5. Dry the outside of the Glass Tube.
6. Re-install the Glass Tube into the FT-2 and press the EXIT soft key.
Note: It is important to use distilled or deionized water only for cleaning.
For more detailed information on cleaning and maintenance, please refer to the FlowTrax FT-2 Flow, Volume Testing device's Operation Manual, specifically the section on cleaning and maintenance.