Answer
Jan 30, 2025 - 09:42 PM
To collect and handle the blood sample for the abionic IVD CAPSULE D-Dimer test, follow these steps:
1. Use a designated trisodium citrate blood tube for venous whole blood collection. Follow the manufacturer's protocol and recommendations for haemostasis tests.
2. Activate the provided abioPIPETTE by turning the plunger clockwise by one quarter.
3. Press on the abioPIPETTE's plunger and maintain the pressure. Insert the abioPIPETTE into the trisodium citrate blood tube.
4. Release the plunger to completely fill the abioPIPETTE with blood and then remove it. 50 µl of blood should be loaded.
It is important to note that the blood sample mixed with the abioMIX reagent should be immediately loaded onto the D-Dimer capsule and the filled capsule should be immediately measured.
For more detailed instructions and information on the test procedures, please refer to the abionic IVD CAPSULE D-Dimer Instructions provided in the manual.
Add New Comment