To install the MT100 Temperature Monitoring Series, follow these steps:
1. Unscrew the top cover of the equipment and remove the insulating strip on the top of the button battery.
2. Tighten the top cover of the equipment, sterilize it, and put it into the incubator to be tested.
3. Pay attention to the location of the equipment. If the bin is relatively large, try to place the equipment close to the outside and keep it away from metal objects as far as possible to avoid signal transmission interference.
4. Ensure that the antenna is placed outward, with the side engraved with SN facing outward to prevent surrounding metal objects from affecting the device's wireless signal.
5. The temperature sensing area is located in the center of the equipment base, and this part should fit the test area during installation.
6. Press the Reset button to reset the whole system. If you want to modify parameters using the serial port or wireless function, do so within the first 30 seconds of resetting the device.
7. Before using the equipment, remove the insulating strip on the top of the button battery to power on the equipment. Retain the side insulating strips for future use.
8. Ensure that the distance between the MT100 and the MT500 or MT1100 receiver is within the transmission range of 3 meters.
Please refer to the "Installation Instructions" section of the METIEC MT100 Temperature Monitoring Series User Manual for more detailed information and illustrations.