Answer

Jan 07, 2025 - 11:43 PM
To calibrate the A&D Weighing FG-150KAL Digital Platform Scale, follow the steps below:
1. Ensure that the scale is clean and free from any objects or debris.
2. Press the "ON/OFF" button to turn on the scale.
3. Wait for the scale to stabilize and display "0.0" on the screen.
4. Press and hold the "CAL" button for approximately 3 seconds until the display shows "CAL".
5. Release the "CAL" button, and the display will show "CAL 0.0".
6. Place a known weight on the scale that is within the scale's capacity. The weight should be evenly distributed on the platform.
7. Press the "CAL" button again, and the display will show "CAL X.XX" (where "X.XX" represents the weight value).
8. Remove the known weight from the scale.
9. The display will show "CAL 0.0" again, indicating that the calibration is complete.
Note: It is recommended to perform calibration regularly to ensure accurate weighing results. Refer to the user manual for more detailed instructions and additional calibration options if needed.