Calibrating the A&D Engineering FG-200KAL Series FG-K Bench Scale is an important step to ensure accurate measurements. Here is a step-by-step guide on how to calibrate the scale, supported by the user manual:
1. Make sure the scale is placed on a stable and level surface.
2. Press the "ON/OFF" button to turn on the scale.
3. Wait for the scale to stabilize and display "0.0" on the screen.
4. Press and hold the "MODE" button for about 2 seconds until the calibration mode is entered. The display will show "CAL".
5. Release the "MODE" button. The display will show the current calibration weight.
6. Place the calibration weight on the scale. The weight should be within the scale's capacity and match the calibration weight shown on the display.
7. Press the "MODE" button to confirm the calibration weight. The display will show "CAL" and the calibration weight.
8. Remove the calibration weight from the scale.
9. The scale will automatically calibrate and display "PASS" if the calibration is successful. If the calibration fails, the display will show "FAIL" and you may need to repeat the calibration process.
10. Press the "MODE" button to exit the calibration mode. The scale is now calibrated and ready for use.
It is important to note that the specific calibration procedure may vary slightly depending on the model and firmware version of your FG-200KAL Series FG-K Bench Scale. For more detailed instructions and illustrations, please refer to the user manual provided in the link you provided.