To calibrate the Health o meter 2500KG Digital Wheelchair Ramp Scale, follow the steps below:
1. Ensure that the scale is on a level surface and there are no objects on the platform.
2. Press and hold the "ON/OFF" button for approximately 3 seconds until the display shows "CAL".
3. Release the "ON/OFF" button. The display will show "0.0".
4. Place a known weight on the scale. The weight should be within the scale's capacity.
5. Press the "ENTER" button. The display will show the weight of the known weight.
6. Remove the known weight from the scale.
7. Press the "ENTER" button again. The display will show "0.0".
8. Calibration is complete.
Note: It is recommended to perform calibration periodically or whenever the scale is moved or if accuracy is in question. The scale should also be calibrated if the batteries are replaced.
For more detailed instructions and illustrations, please refer to the user guide provided in the link: [Health o meter 2500KG Digital Wheelchair Ramp Scale User Guide](https://manuals.plus/wp-content/uploa...)