Answer
Jan 05, 2025 - 10:38 PM
To calibrate the Steinberg SBS-TW-3000/100G Digital Table Scale, please follow the steps below:
1. Switch on the scale by pressing the "ON / OFF" button.
2. Press and hold the "M" button until the display shows the calibration value (3000 in this case).
3. Release the "M" button and press it again. The display will show "CAL" and a flashing standard load value in grams (3000 in this case).
4. Place the indicated standard load value (3000 grams) on the scale's weighing platform and wait for the scale to stabilize.
5. Once the scale has stabilized, the display will show another flashing standard load value in grams (3000 in this case). Place this second standard load value on the weighing platform.
6. When the calibration is completed successfully, a flashing "PASS" message will be displayed on the screen, and the scale will turn off.
Please note that it is important to use the correct standard load value (3000 grams) for calibration. If you encounter any issues or have further questions, it is recommended to contact your local representative or customer service for assistance.
Add New Comment