To calibrate your AWS pocket scale, follow these steps:
1. Place the required calibration weight on the scale's weighing platform.
2. Note the displayed value on the scale.
3. If the displayed value is not within acceptable tolerance, adjustment is required.
4. With the scale turned on, press and hold the Mode (M) key until the display shows "CAL."
5. Press the Mode (M) key again. The display will flash "CAL" followed by the required calibration weight.
6. Carefully place the required calibration weight on the center of the platform.
7. Once the weight has stabilized, the scale will show "PASS" and then return to normal weighing mode. Calibration is complete!
Please refer to the "Calibration" section in the user manual for more detailed instructions and the acceptable tolerance range for calibration.