To calibrate the DigiPas DWL90Pro 2-AXIS SMART CUBE DIGITAL LEVEL, follow these step-by-step instructions:
1. Press and hold the 'Alt. Zero' buttons while turning on the device. The screen will display "CAL 1" flashing.
2. Place the device on a flat and clean surface. Press the power button once, and the screen will display a countdown. Wait until the countdown reaches "1" and shows "CAL 2".
3. Rotate the device 180 degrees parallel to the surface. Press the power button again, and the screen will display a countdown. Wait until the countdown reaches "1" and shows "CAL 3".
4. Position the device vertically (-90°) on a vertical surface. Press the power button again, and the screen will display a countdown. Wait until the countdown reaches "1" and shows "CAL 4".
5. Rotate the device 180 degrees parallel vertically to the surface (+90°) as shown in Figure 4. Press the power button again, and the screen will display a countdown. Wait until the countdown reaches "1".
6. Calibration is now complete.
Note: It is important to ensure that the device is kept still during calibration for accurate results.
For more information and visual demonstration on how to calibrate the device, you can refer to the video demonstration available on the DigiPas website at www.digipas.com.
Please note that the DigiPas DWL90Pro 2-AXIS SMART CUBE DIGITAL LEVEL is calibrated in the factory for maximum accuracy. Users can re-calibrate the device when necessary.