Answer
Dec 20, 2024 - 05:47 AM
To calibrate the Neoteck NTK033 Digital Angle Finder, follow these steps:
1. Turn off the equipment and place it on a flat and smooth solid surface.
2. In the shutdown status, press and hold the HOLD button first, then press the ON/OFF button until the LCD display reads "CAL".
3. Keep the equipment still for a period of time and then press the HOLD button. The display will read "CAL1".
4. In the same location, rotate the equipment 180° and then press the HOLD button. The display will read "CAL2".
5. After the "CAL2" disappears, the calibration is completed, and the LCD will show the current angle in high-accuracy display.
Note:
- The calibration must be performed on a flat and smooth surface with a tilt of no more than 5°. If it exceeds 5°, the display will read "Err", indicating that calibration cannot be performed.
- During calibration, the level should not be beveled or moved.
- After completing calibration, the equipment will automatically replace the original factory data with the user's data, so caution must be exercised when using this function.
Please refer to the Neoteck NTK033 Digital Angle Finder User Manual for more detailed instructions and illustrations.